Solid Autoignition Temperature Tester
This instrument is designed to determine the relative autoignition temperature of solid substances. The sample is loaded into a 20 mm standard cubic wire-mesh basket and heated at a constant rate in a programmable temperature-controlled oven to precisely observe the critical temperature at which the substance undergoes spontaneous ignition due to self-heating oxidation. Applicable to industrial chemicals, pesticide technical materials, and solid powders, it provides critical technical data for fire hazard classification of chemicals, process safety temperature specifications, and identification of self-ignition characteristics in hazardous waste.
Test Parameters
Relative Autoignition Temperature
Application
Chemicals, Pesticides, Environmental Protection
Features
Embedded processor, Windows CE operating system.
8-inch LCD touchscreen displaying real-time test status.
Supports Standard and Custom modes (custom mode allows saving of exothermic peaks, etc.), balancing standard compliance and flexibility.
Integrates high-temperature insulation, precision temperature control, and advanced detection technologies.
Electrically heated crucible furnace with natural convection for uniform sample heating.
Furnace chamber lined with vacuum-insulated ceramic fiber for excellent thermal insulation.
Wire-mesh cube suspended from furnace cover bottom for safe and convenient sample loading.
Automatic heating control, temperature control, and combustion detection.
Includes push rod for cleaning central orifice of inner liner.
Includes collection tray for samples falling from wire-mesh cage or cleared by push rod.
Includes cooling stand for furnace cover assembly post-test.
Thermocouple sensors for rapid and accurate temperature response.
Real-time display of time-temperature profiles with access to raw data.
Fully automated test process requiring no manual intervention; supports parameter setting, data storage, and report viewing.
Classical PID control algorithm to minimize temperature overshoot caused by thermal inertia, ensuring constant heating rate.
Stainless steel crucible with high corrosion resistance ensures temperature uniformity in high-aspect-ratio furnace; natural convection ensures uniform sample heating.
High-precision thermocouples for accurate monitoring of chamber and sample temperatures.
Specifications
| Operating Environment | -5–45 °C, <95% RH |
| Test Object | Solid chemical products |
| Temperature Control Range | RT–650 °C |
| Furnace Volume | ≥2 L |
| Furnace Material | Stainless steel, ceramic fiber |
| Standard Heating Rate | 0.5 °C/min |
| Custom Heating Rate | 0.1–3.0 °C/min |
| Temperature Accuracy | ±0.004|t| °C |
| Temperature Measurement Range | -40–1000 °C |
| Temperature Measurement Resolution | 0.1 °C |
| Temperature Measurement Accuracy | ±2.5 °C (<300 °C); ±0.75% of reading (≥300 °C) |
| Heating Chamber Capacity | 2 L |
| Timing Range | 0–96 h |
| Timing Accuracy | ±0.01 s |
| Mesh Cage Dimensions | 20 mm side length, 0.045 mm aperture |
| Instrument Dimensions (L × W × H) | 840 mm × 430 mm × 750 mm |